TOC Analysis of Pure and Ultra Pure Water
- Low sample volume / lower cost and waste
- 9 months UV lamp life guarantee / 50% longer life
- Continuous measurement (every 2 seconds) / fast reaction
- Offline measurement / grab samples can be measured
- Up to 3 sample inlets / very efficient
- Remote service / online service diagnosis
Overview
Designed for Total Organic Carbon monitoring online and offline
The miniTOC instrument uses the classical flow-through concept for online or offline TOC monitoring of pure and ultra pure water.
The efficient UV oxidation of organic compounds by the “direct surface” principle in combination with conductivity detection guarantees the fast monitoring of values.
Calibration and SST of the miniTOC is easy and can be done by the user.
Online operation is interrupted briefly for the system checks to be performed.
The operator simply has to connect the inlet with the calibration standard solutions and the diluting water.
The software enables the user to perform calibration and SST independently, and the programmable sequence ensures the automatic procedure and data evaluation of the test.
Verification of the TOC results according to the requirements of USP and Ph.Eur is realised by the periodic performance of system suitability tests.
Main Technical Specifications of the System
- Method & detection: UV-Oxidation & conductivity
- Sample volume flow: ca. 14 mL/min
- Total Organic Carbon (TOC) Range: 0.5 to 1000 ppb
- Related norms: USP <643>, EP 2.2.44, DIN EN 1484
Technical Specifications
- Measurement range: 0.5 to 1000 ppb
- Ports: 1 for sample, 1 for SST (basic) | 1 for sample, 3 for SST (pro)
- Time of measurement: 2 sec (online mode) | 1 min (offline mode)
- Dimensions, weight, power: 300 x 200 x 500 mm, 14 kg, 110 - 230 V, 50 Hz, 60 Watt
- Interface ports: Analog, according to Namur NE43
- Sample volume flow: Approx. 14 mL/min
- Ambient parameters: Relative humidity: 10 to 95%, Temperature: 10 to 45°C
- Related norms: USP <643>, EP 2.2.44, DIN EN 1484
- Method & detection: UV oxidation (partial) & conductivity
- Detection accuracy: ~0.5 ppb or 1% (whichever is greater)
Medium Requirements
- Type of sample medium: Pure or ultra pure water
- Medium conductivity: < 2 µS/cm (upgradable to < 10 µS/cm)
- Inlet pressure: < 1 bar
- Sample temperature: < 50°C