Aquinity² P35 / P70 Ultra Pure Water System
Key Features
- Produces Pure Water and Ultra Pure Water
- Compact housing
- Big touchscreen
- Remote service module
- Menus and wizards easy to use and mistake avoiding
- Fixed and flexible dispenser
- Easy access for cartridge exchange
Overview
Designed for the production of pure water and ultra pure water out of tap water.
The Aquinity² P versions produce deionised water (< 1 µS/cm/ > 1 MOhm x cm) by reverse osmosis and conditioning module into the external reservoir (35 liter or 70 liter).
The water can be tapped directly (e.g., for glassware cleaning by hand) or by connecting the tank to a laboratory dishwasher.
Out of the storage tank, ultra pure water is produced and can then be tapped with a flexible dispenser.
The Aquinity² water purification system is available in different configurations (Reagent, Analytical & LifeScience) to meet the specific requirements of ultra pure water for various applications:
- Analytical: Includes UV lamp for low TOC value.
- Life Science: Includes UV lamp and Ultrafiltration Module for low TOC and low endotoxin values.
An optional integrated TOC sensor allows the user to monitor the TOC content permanently online.
The Aquinity² can be configured as a bench-top or bench-integrated system with an external panel, or as a wall-mounted system (with inclined display). The tank can be installed at the side, above the system, or underneath the bench.
Upon request, the tanks can be connected in series to reach a larger reservoir volume.
Main Technical Specifications
- Resistivity/Conductivity:
- > 1 MOhm x cm | < 1 µS/cm; Type II
- 18.2 MOhm x cm | 0.055 µS/cm; Type I
- Production rate: Max. 10 L/h, optionally 20 L/h
- Total Organic Carbon (TOC):
- < 10 ppb (Reagent)
- < 5 ppb (Life Science)
- < 3 ppb (Analytical)
Technical Specifications
- Pure water + ultrapure water quality: < 2 µS/cm; Type II + 0.055 µS/cm; Type I
- Resistivity: > 0.5 MegaOhm; Type II + 18.2 MegaOhm/cm; Type I
- Total Organic Carbon (TOC): < 10 ppb Reagent, < 5 ppb Life Science, < 3 ppb Analytical
- Dispensing flow rate: 2.0 L/min (1.5 L/min Life Science)
- Production rate: 10 L/h, optionally 20 L/h
- Bacteria: < 1 cfu/mL
- Particulate: > 0.2 µm less than 1 particulate/mL
- Pyrogen (endotoxins): < 0.001 EU/mL*
- RNAse: < 1 pg/mL*
- DNAse: < 5 pg/mL*
Feed Water Requirements
- Type of feed water: Tap water
- Feed water conductivity: < 1400 µS/cm
- Inlet pressure: 1.5 to 6 bar
- Free chlorine: < 0.1 mg/L
- Silt Density Index (SDI): < 3
- pH: 3 to 9
- Temperature: 5 to 25 °C